Stanford SR715 LCR Meter

multi-frequency passive components measuring instrument (Distributed only in Germany.)

Specifications

Measurement modesAuto, R+Q, L+Q, C+D, C+R
Equivalent circuitSeries or Parallel
Parameters displayedValue, Deviation, % Deviation or Bin Number.
Deviation and % Deviation are calculated relative to a stored value.
Averaging2 to 10 Measurements
Display Range
R+Q
R0.0001 Ohm - 2000 MOhm
Q0.00001 - 50
L+Q
L0.0001 mH - 99999 H
Q0.00001 - 50
C+D
C0.0001 pF - 99999 mF
D0.00001 - 10
C+R
C0.0001 pF - 99999 mF
R0.00001 Ohm - 99999 kOhm
Test conditions
Test frequency Fixed frequencies at 100 Hz, 120 Hz,1 kHz, 10 kHz (SR715 and SR720) 100 kHz (SR720 only).
Frequency accuracy ±100 ppm.
Drive voltagePreset levels: 0.10, 0.25, and 1.0 Vrms.
Vernier: 0.1 to 1.0 Vrms with 50 mV resolution.
Drive levels accurarcy±2%.
Measurement rateSlow, Medium, Fast: 2, 10, or 20 measurements per second at test frequencies of 1 kHz and above and about 0.6, 2. 4, or 6 measurements per second at 100 Hz and 120 Hz.
RangingAuto or Manual
TriggeringContinuous, Manual, or Remote over RS-232, GPIB or Handler Interface
Bias voltageInternal: 2.0 VDC ±2 %, external: 0 to +40 VDC, fused @ 250 mA
Accuracy
Basic accuracy SR715: 0.20 %, SR720: 0.05 % (see graphs in datasheet for important details!)
Features
Fixture4-Wire Kelvin fixture for radial leaded parts with adapters for axial leaded parts.
ProtectionProtected up to 1 Joule of stored energy (for charged capacitors). Fused at 0.25 A output current for biased measurement.
Zeroing Open and short circuit compensation.
Compensation limitsShort: R <20 Ohm, Z <50 Ohm, Open: Z >10 kOhm
BinningUp to 8 pass bins, QDR and general fail bins, defined from the front panel or over the computer interfaces. Binning setup may be stored in non-volatile memory.
Store and recallStores 9 Complete Instrument Setups.
RS-232 interfaceAll instrument functions can be controlled or read over the interface.
GPIB and handlerIEEE-488 interface optional, Handler interface is a 25 pin connector, positive logic for binning and control
General
Operating conditions0 to 50 °C, <85 % relative humidity.
Power 20 Watts, 100/120/220/240 VAC, 50 or 60 Hz.
Dimensions 13.5 " W x 14 " L x 4 " H
Weight10 lbs
Warranty One year parts and labor on any defects in material or workmanship.