R&S®EZ-25 150 kHz Highpass Filter

For conducted emission measurements in the presence of high longwave mains disturbance signals

Key Facts
Picture R&S®EZ-25 150 kHz Highpass Filter
  • Conducted emission measurements in line with EN 50065 Part 1
  • Pass frequency range from 150 kHz to 30 MHz
  • Very steep slope in line with CISPR 16-1-1
  • Suitable for any CISPR measuring receiver
  • Relative attenuation >50 dB below 130 kHz
  • Built-in 10 dB attenuator for exact 50 Ω termination of the LISN
  • High pulse energy capability (50 mWs)
  • Calibration data supplied