| Number of test ports | two or four | ||
|
Frequency range
|
300 kHz to 8 GHz ( |
||
| 10 MHz to 24 GHz ( |
|||
| 10 MHz to 40 GHz ( |
|||
| 10 MHz to 50 GHz ( |
|||
| 10 MHz to 67 GHz ( |
|||
| 10 MHz to 80 GHz ( |
|||
| 10 MHz to 110 GHz ( |
|||
| Measurement time per test point | |||
| At 1 MHz IF bandwidth | <3.5 µs | ||
| At 5 MHz IF bandwidth (with
|
<2.0 µs | ||
| Sampling time per test point | |||
| At 1 MHz IF bandwidth | 1.8 µs | ||
| At 5 MHz IF bandwidth (with |
430 ns | ||
| Measurement time (201 test points) | 4.5 ms | ||
| Data transfer time (201 test points) | |||
|
Via IEC/IEEE bus
|
<2.9 ms
|
||
|
Via VX11 over 100 Mbit/s LAN
|
<1.3 ms
|
||
|
Via RSIB over 100 Mbit/s LAN
|
<0.7 ms
|
||
| Switching time | |||
| Between channels | <1 ms | ||
| Between instrument setups | <10 ms | ||
| Dynamic range at 10 Hz measurement bandwidth | |||
| Between test ports | typ. >140 dB | ||
| With direct receiver access | typ. >145 dB | ||
| Dynamic range at 1 Hz measurement bandwidth with direct receiver access |
typ. >150 dB |
||
| Output power at test port |
>13 dBm, typ. 15 dBm
|
||
| Power sweep range |
>50 dB, typ. 60 dB
|
||
| IF bandwidth |
|
||
| Standard | 1 Hz to 1 MHz | ||
| With |
5 MHz (10 MHz with restrictions) | ||
| With |
10 MHz (30 MHz with restrictions) | ||
| Number of channels, diagrams, traces |
>100 each
|
||
| Number of test points per trace |
1 to 60001
|
||
| Operating system |
Windows XP Embedded
|
||
Instrument concept
The

Direct generator and receiver access options (blue); generator and receiver step attenuator options (red)
The above design features make the
- Use of the second source as an LO for fast mixer measurements
- Generation of a swept two-tone signal for fast intermodulation measurements versus frequency or power
- True differential measurements on balanced DUTs
- Parallel signal output and measurement at the test ports for measuring two or more DUTs simultaneously.
- Direct generator / receiver access (option) enabling the following, for example:
- Loop-in of preamplifiers into generator paths
- Loop-in of attenuators into the receiver paths
- Connection of external test setups, e.g. power amplifiers, or use of analyzer as a multichannel receiver, e.g. for antenna measurements