R&S®ZVA-K6 True Differential Measurements

Features

The R&S®ZVA in the true differential mode offers the following features:

  • Easy to configure
  • The steps required to configure and calibrate true differential measurements are identical to those for virtual differential measurements. You can switch between the two modes simply at a mouse click.

  • Simultaneous measurement and display of mixed-mode S-parameters in the true differential and virtual differential modes
  • Switching between the true differential and the virtual differential mode takes no more than a mouse click. When you have configured two measurement channels, you can measure a DUT in both modes simultaneously.

  • System-error-corrected mixed-mode wave quantities (power levels) and S-parameters
  • Using a conventional four-port calibration in conjunction with a special, patented correction algorithm, you can generate and measure system-error-corrected mixed-mode wave quantities (power levels) and S-parameters of the reference plane.

  • Characterization of the DUT´s using real-world signals
  • The option allows the generation of imperfectly balanced signals, i.e. signals that contain common-mode as well as differential-mode (normal-mode) components, or even purely common-mode components. The option also supports true phase imbalance and true amplitude imbalance sweeps of the two stimulus signals.

  • High measurement speed
  • The high measurement speed results in short test and tuning times.

  • High phase accuracy up to 50 GHz
  • A new, patented technique for establishing the phase relationship between the two source signals (phase locking) ensures small phase deviation of typically below 1° over the entire frequency range with minimum time drift.