R&S®FS-K15 VOR/ILS Measurement Demodulator
for R&S®FSMR /R&S®FSU /R&S®FSQ - Analysis for VOR and ILS signals
Key Facts
| Frequency range: | 107 MHz to 120 MHz and 319 MHz to 341 MHz specified; can be used in the frequency range of the base unit | |
| Level range: | -60 dBm to +30 dBm | |
| Measurement functions VOR: | VOR phase, modulation depth and frequency for 30 Hz, subcarrier, identifier, 9.96 kHz subcarrier deviation, modulation depth and frequency of the subcarrier, THD; phase between 90 Hz and 150 Hz signal | |
| VOR phase: | 0° to 360°, 0.1° resolution | |
| Measurement uncertainty VOR phase: | 0.003° | |
| Measurement functions ILS: | DDM, SDM, modulation depth and frequency at 90 Hz/150 Hz/identifier, THD | |
| Measurement range DDM: | 0 DDM to 0.4 DDM | |
| Measurement uncertainty DDM: | 0.0002 DDM + 1 % of reading | |