Option 76 3 GHZ Tracking generator

Key Facts

Option 76: Tracking generator up to 3 GHz
Frequency range 100 KHz to 3 GHz
Frequency offset  
  Range 0 to 1 GHz  
  Resolution 1 kHz  
  Accuracy +/- 200 Hz  
Output level range 0 to -60 dBm 0.5 dB step
Output level accuracy +/- 0.5 dB 0.5 dB step
Output level flatness   Reference signal level: -10 dBm, frequency: 20 MHz
    +/- 1.0 dB (1 MHz to 1 GHz)  
    +/- 1.5 dB (100 KHz to 3 GHz)  
Output level switching uncertainly   Reference level: -10 dBm
    +/- 1.0 dB (10 MHz to 1 GHz)  
    +/- 2.0 dB (10 MHz to 2.6 GHz)  
    +/- 3.0 dB (10 MHz to 3 GHz) Freq. Offset OFF
    +/- 5.0 dB (10 MHz to 3 GHz) Freq. Offset ON
Spurious output   Output level: -10 dBm
  Harmonic <= -20 dBc  
  Non-harmonic <= -25 dBc Freq. Offset OFF
TG leakage <= -80 dBm Input attenuator: 0 dB
Output impedence 50 ohm (nominal)  
  VSWR <= 2 (characteristic value) Output level <= -10 dBm
Maximum allowable input + 10 dBm, +/- 10 V